7.2 Driver or sensor coil fault

Typical inductance and resistance values

7.2.1 OPTIMASS 1000

The specified values are for guidance only.

• Driver = black and grey
• Sensor A = white and yellow
• Sensor B = green and violet
• Pt500 = red and blue (530…550 Ω) at ambient temperature
• Strain values measuring tube = 420...560 Ω
• Resistance values outside of the ranges specified above may indicate a circuit error. The

device may be in "start-up" mode or indicate measuring errors.

• All circuits should be isolated with respect to earth (device housing) and each other >20 MΩ.
• An exciter/sensor coil short circuit may cause the meter to be in "start-up" mode.

Size

Resistance (Ohm)

Driver

Sensor A/B

15

220

78

25

220

64

40

163

78

50

163

64

15-Ex

220

78

25-Ex

220

64

40-Ex

94

78

50-Ex

94

64

INFORMATION!
Failure of two or more of the upper circuits may indicate a measuring tube error. There may be
product in the housing. If this is the case, depressurise

depressurise

depressurise

depressurise the process line and immediately remove

the measuring device from the process line.