6.4.4 Set free units

C6.8 HART

C6.8

HART

C6.8.1

HART

Switch HART

®

communication on/off:

Select: HART on (HART

®

activated) current = 4…20 mA / HART off (HART

®

not activated) current = 0…20 mA

C6.8.2

Address

Set address for HART

®

operation:

Select: 00 (point-to-point operation, current output has normal function,

current = 4…20 mA) / 01...15 (multidrop operation, current output has a

constant setting of 4 mA)

C6.8.3

Message

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C6.8.4

Description

Set required text:

A…Z ; a…z ; 0…9 ; / - + , . *

C6.9 Quick Setup

C6.9

Quick setup

Activate quick access in Quick Setup menu:

Select: yes (switched on) / no (switched off)

C6.9.1

Reset Totaliser 1

Reset Totaliser 1 in Quick Setup menu?

Select: yes (activated) / no (switched off)

C6.9.2

Reset Totaliser 2

Reset Totaliser 2 in Quick Setup menu?

Select: yes (activated) / no (switched off)

C6.9.3

Reset Totaliser 3

Reset Totaliser 3 in Quick Setup menu?

Select: yes (activated) / no (switched off)

Free units

Sequences to set texts and factors

Texts

Texts

Texts

Texts

Volume flow, mass flow and density:

3 digits before and after the slash

xxx/xxx (max. 3 digits before / after the slash)

Volume, mass:

xxx (max. 3 digits)

Permissible characters:

A…Z ; a…z ; 0…9 ; / - + , .*; @ $ % ~ () [] _

Conversion factors

Conversion factors

Conversion factors

Conversion factors

Desired unit

= [unit see above] * conversion factor

Conversion factor

Max. 9 digits

Shift decimal point:

↑ to the left and ↓ to the right