Operation – KROHNE MFC 300 EN User Manual
Page 96

6
OPERATION
96
MFC 300
www.krohne.com
02/2012 - 4000498602 - MA MFC 300 R03 en
6.4.4 Set free units
C6.8 HART
C6.8
HART
C6.8.1
HART
Switch HART
®
communication on/off:
Select: HART on (HART
®
activated) current = 4…20 mA / HART off (HART
®
not activated) current = 0…20 mA
C6.8.2
Address
Set address for HART
®
operation:
Select: 00 (point-to-point operation, current output has normal function,
current = 4…20 mA) / 01...15 (multidrop operation, current output has a
constant setting of 4 mA)
C6.8.3
Message
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C6.8.4
Description
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C6.9 Quick Setup
C6.9
Quick setup
Activate quick access in Quick Setup menu:
Select: yes (switched on) / no (switched off)
C6.9.1
Reset Totaliser 1
Reset Totaliser 1 in Quick Setup menu?
Select: yes (activated) / no (switched off)
C6.9.2
Reset Totaliser 2
Reset Totaliser 2 in Quick Setup menu?
Select: yes (activated) / no (switched off)
C6.9.3
Reset Totaliser 3
Reset Totaliser 3 in Quick Setup menu?
Select: yes (activated) / no (switched off)
Free units
Sequences to set texts and factors
Texts
Texts
Texts
Texts
Volume flow, mass flow and density:
3 digits before and after the slash
xxx/xxx (max. 3 digits before / after the slash)
Volume, mass:
xxx (max. 3 digits)
Permissible characters:
A…Z ; a…z ; 0…9 ; / - + , .*; @ $ % ~ () [] _
Conversion factors
Conversion factors
Conversion factors
Conversion factors
Desired unit
= [unit see above] * conversion factor
Conversion factor
Max. 9 digits
Shift decimal point:
↑ to the left and ↓ to the right
.book Page 96 Monday, February 6, 2012 11:01 AM