Electrical connections – KROHNE MFC 300 EN User Manual

Page 63

Advertising
background image

ELECTRICAL CONNECTIONS

4

63

MFC 300

www.krohne.com

02/2012 - 4000498602 - MA MFC 300 R03 en

Pulse and frequency output passive P

N

NAMUR, Ex i I/Os

• Connection acc. to EN 60947-5-6
• open:

I

nom

= 0.43 mA

closed:
I

nom

= 4.5 mA

• X designates the connection terminals B or D, depending on the version of the signal

converter.

DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.

INFORMATION!
•

For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from
electrical interferences (EMC).

•

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions:

Compact and field housing versions: Shield connected via the cable terminals in the terminal
compartment.
Wall-mounted version:

Wall-mounted version:

Wall-mounted version:

Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
(insulation to DIN 46245) in the terminal compartment.

•

Any connection polarity.

Figure 4-30: Pulse and frequency output passive P

N

as per NAMUR EN 60947-5-6 Exi

.book Page 63 Monday, February 6, 2012 11:01 AM