KROHNE IFC 010 Converter EN User Manual

Page 14

Advertising
background image

3/1

3 Start-up

•

Before powering the system, please check that it has been correctly installed according
to Sect. 1 and 2.

•

The flowmeter is delivered ready for operational use. All operating data have
been factory set in accordance with your specifications.
Please refer to Sect. 3.2 “factory settings”.

•

Power the unit, and the flowmeter will immediately start process flow
measurement.

Basic version, signal converter IFC 010 / B
•

A light emitting diode (LED) under the cover of the electronic section shows the
measurement status.
LED flashing . . .

green:

measurement correct, everything all right.

green / red: momentary overdriving of outputs

and/or A/D converter.

red:

fatal error, parameter error or hardware fault,
please consult factory.

•

Refer to Sect. 6.1 for operator control of the “basic version”.

Display version, signal converter IFC 010 / D
•

When powered, the display shows in succession: START UP and READY.
This is followed by display of the current flow rate and/or the current totalizer count
on either a continuous or alternating basis, depending on the setting under Fct. 1.04.

•

Refer to Sect. 4 and 5 for operator control of the “display version”.

3.1 Power- on and measurement