KROHNE IFC 010 Converter EN User Manual
Page 29

5/4
Current output I 5.6
Fct. 1.05 CURRENT I
Press
→
key.
→
FUNCT. I = select function for current output,
press
→
key
•
OFF
(switched off, no function)
•
1 DIR.
(1 flow direction)
•
2 DIR.
(2 flow directions, F/R mode, forward/reverse)
Select using
↑
or
↓
key
.
Transfer to subfunction “RANGE I” with
↵
key.
Exception: When “OFF” selected, return to Fct. 1.05 CURRENT I.
→
RANGE I = select measuring range,
press
→
key
•
0 - 20 mA
fixed ranges
•
4 - 20 mA
Press
→
key to transfer to number setting.
Select with key
↑
or
↓
.
Press key
↵
to transfer to subfunction “I ERROR”.
→
I ERROR = set error value,
press
→
key
•
0
.0
mA
•
3.6 mA
(only possible if range ”4-20 mA” selected)
•
22
.6
mA
Select using key
↑
or
↓
.
Press
→
key to transfer to number setting.
Press key
↵
to return to Fct. 1.05 CURRENT I.
Please refer to Sect. 3.2 “Factory settings”.
Refer to Sect. 2.3 for connection diagrams, and to Sect. 5.14 for characteristics.