4/8

Fct.

Text

Description and settings

3.00

INSTALL.

Installation menu

3.01

LANGUAGE

Select language for display texts
•
GB / USA (English)

• F (French)

• D (German)

• others on request

Press

↵

key to return to Fct. 3.01

“LANGUAGE”.

3.02

FLOWMETER

Set data for primary head

→

DIAMETER

Select size from meter size table
•
DN 10 - 1000 mm equivalent to 3/8 - 40 inch

Select with

↑

or

↓

key.

Press

↵

key to transfer to subfunction “FULL SCALE ”.

→

FULL SCALE

Full-scale range for flow Q

100%

To set, refer to Fct. 1.01 “FULL SCALE” above.

Press

↵

key to transfer to subfunction “GKL VALUE”.

→

VALUE P

Pulse value (Fct. 1.06

“VALUE P”

) has been changed.

With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Check new values!

→

GKL VALUE

Set primary constant GKL
see primary head nameplate.
Range:

• 1.0000 - 9.9999

Press

↵

key to transfer to subfunction

“FIELD. FREQ.”.

→

FIELD FREQ.

Magnetic field frequency
Values: 1/6 or 1/18 of power frequency, see nameplate.

Press

↵

key to transfer to subfunction “FLOW DIR.”;

DC units only: to transfer to subfunction

“LINE FREQ.”.

→

LINE FREQ.

Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:

50 Hz

and

60 Hz

Press

↵

key to transfer to subfunction “FLOW DIR.”.

→

FLOW DIR.

Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
• + DIR.

• – DIR.

Select using

↑

or

↓

key.

Press

↵

key to return to Fct. 3.02 “FLOWMETER ”.

3.03

ZERO SET

Zero calibration
Note:

Carry out only at “0” flow and with completely filled
measuring tube!

Precautionary query
• CALIB. NO

Press

↵

key to return to Fct. 3.03 “ZERO SET”.

• CALIB. YES

Press

↵

key to start calibration.

Duration approx. 15 –90 seconds, current flowrate
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).

A “WARNING” sign appears when flowrate “>

0

“;

acknowledge by pressing

↵

key.

• STORE NO (do

not store new zero value)

• STORE YES (store new zero value)

Press

↵

key to return to Fct. 3.03 “ZERO SET”.

3.04

ENTRY CODE

Entry code required to enter setting mode?
•
NO

(= entry with

→

only)

• YES (= entry with

→

and Code 1:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

)

Press

↵

to return to Fct. 3.04 “ENTRY CODE”.