KROHNE IFC 010 Converter EN User Manual

Page 45

Advertising
background image

7/6

Part 2

Signal converter IFC 010 D (D = display version) and

(cont’d)

Signal converter IFC 010 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.1)

Group P

Fault / Symptom

Cause

Remedial action

P 1

Totalizer connected but does

Incorrect connection/polarity

Connect properly,

not count any pulses

see Sect. 2.3

Totalizer or external voltage

Check output (see Sect. 7.2)

source defective.

with new totalizer:
Test ok, check connection
cables and previous
totalizers and external voltage
source, and replace if
necessary.
Test faulty, pulse output
defective, replace signal
converter (see Sect. 8.4)
or contact Krohne Service.

Current output is external

Check connection and cables,

voltage source, short circuit

see Sect. 2.3. Voltage between

or current / pulse output

I+ and I

approx. 15 V.

defective

Switch off device, eliminate
short circuit, switch device
on again. If no function,
then current or pulse
output defective. Replace
signal converter (see Sect.
8.4) or contact Krohne
Service.

Pulse output is deactivated,

Switch on under Fct. 1.06.

see Fct. 1.06.

P 2

Unsteady pulse rate

– Process product

Increase or switch on

conductivity too low,

time constant,

particles/air inclusions

see Sect. 6.1, or contact

too large or inhomogeneous

Krohne Service.

– Pulsating flow
– Time constant too low

or switched off

P 3

Pulse rate too high or

Incorrect setting for

Change setting under Fct. 1.06.

too low.

pulse output.

Group S

Fault / Symptom

Cause

Remedial action

S 1

No function

Incorrect connection/polarity

Connect properly,

of status display

see Sect. 2.3.

Status display or output

Set status output under Fct. 1.07

defective or external

to “F/R INDIC.“ (flow direction)

voltage source not

and check (see Sect. 7.2)

supplying voltage.

with new status display:
Test ok, check previous
status display and external
voltage source, and replace
if necessary.
Test faulty, status output
defective, replace signal
converter (see Sect. 8.4) or
contact Krohne Service.

Group D/I/P/S

Fault / Symptom

Cause

Remedial action

D / I / P / S 1

Unsteady display and

– Process product

Increase time constant,

outputs

conductivity too low,

see Sect. 6.1, or contact

particles/air inclusions

Krohne Service.

too large or inhomogeneous

– Pulsating flow
– Time constant too low

D / I / P / S 2

No display and no

Power OFF

Switch power on.

function of outputs

Check power fuse(s)

Replace if defective,

F1 (F1 and F2 for DC).

see Sect. 8.2.