KROHNE IFC 010 Converter EN User Manual

Page 27

Advertising
background image

5/2

Low-flow cutoff 5.3

Display 5.4

Fct. 1.04 DISPLAY
Press

→

key.

→

DISP. FLOW = select unit for display of flowrate,

press

→

key

•

NO DISP.

(no display)

•

m3/hr

(cubic metres per hour)

•

Liter/Sec

(litres per second)

•

US.Gal/min

(US gallons per minute)

•

user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.14

•

PERCENT

(percentage display)

•

BARGRAPH

(numerical value and bar graph display in %)

Select with

↑

or

↓

key.

Press

↵

key to transfer to subfunction “DISP. TOTAL”.

→

DISP. TOTAL = select unit for totalizer display,

press

→

key

•

NO DISP.

(no display)

•

OFF

(internal totalizer switched off)

• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (

ΣΣ

) • ALL (sequential)

Select with

↑

or

↓

key.

Transfer to totalizer unit setting using

↵

key.

•

m3

(cubic metres)

•

Liter

(litres)

•

US.Gal

(US gallons)

•

user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.14

Select with

↑

or

↓

key.

Transfer to totalizer format setting using

→

key.

Continuation see next page

Fct. 1.03 L.F.CUTOFF
Press

→

key.

Choice
•

OFF

(fixed trip points:

ON = 0.1 % / OFF = 0.2 %,

for 100 or 1000 Hz, see Fct. 1.06, 1% or 2%)

•

PERCENT

(variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)

Select with

↑

or

↓

key.

Transfer to number setting using

→

key (only if “PERCENT” selected).

1st number (cursor) flashes.

Setting the numerical value when “PERCENT” selected
•

01 to 19

(cutoff “on” value, left of hyphen)

•

02 to 20

(cutoff “off” value, right of hyphen)

Change flashing number (cursor) with the

↑

or

↓

key.

Shift cursor 1 place to right using

→

key.

Press

↵

key to return to Fct. 1.03 L.F.CUTOFF.

Note: The cutoff “off” value must be greater than the cutoff “on” value.