KROHNE IFC 010 Converter EN User Manual

Page 27

Advertising
background image

5/2

Low-flow cutoff 5.3

Display 5.4

Fct. 1.04 DISPLAY
Press

key.

DISP. FLOW = select unit for display of flowrate,

press

key

NO DISP.

(no display)

m3/hr

(cubic metres per hour)

Liter/Sec

(litres per second)

US.Gal/min

(US gallons per minute)

user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.14

PERCENT

(percentage display)

BARGRAPH

(numerical value and bar graph display in %)

Select with

or

key.

Press

key to transfer to subfunction “DISP. TOTAL”.

DISP. TOTAL = select unit for totalizer display,

press

key

NO DISP.

(no display)

OFF

(internal totalizer switched off)

• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (

ΣΣ

) • ALL (sequential)

Select with

or

key.

Transfer to totalizer unit setting using

key.

m3

(cubic metres)

Liter

(litres)

US.Gal

(US gallons)

user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.14

Select with

or

key.

Transfer to totalizer format setting using

key.

Continuation see next page

Fct. 1.03 L.F.CUTOFF
Press

key.

Choice

OFF

(fixed trip points:

ON = 0.1 % / OFF = 0.2 %,

for 100 or 1000 Hz, see Fct. 1.06, 1% or 2%)

PERCENT

(variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)

Select with

or

key.

Transfer to number setting using

key (only if “PERCENT” selected).

1st number (cursor) flashes.

Setting the numerical value when “PERCENT” selected

01 to 19

(cutoff “on” value, left of hyphen)

02 to 20

(cutoff “off” value, right of hyphen)

Change flashing number (cursor) with the

or

key.

Shift cursor 1 place to right using

key.

Press

key to return to Fct. 1.03 L.F.CUTOFF.

Note: The cutoff “off” value must be greater than the cutoff “on” value.

Advertising