Martel Electronics 235 User Manual

Page 22

Advertising
background image

If the input to the instrument cannot be generated by the calibrator,
the calibrator will be configured for Manual output. An example might
be a magnetic flow meter that requires a manufacturer specific device
to simulate a flow into its input.

There is an additional feature associated with Manual outputs. The
Calibration Database can be configured to require that the user
generate the exact value specified, or that the user generate
approximately the values specified and then enter the exact value that
was set.

If the calibration procedure specifies exact values, the calibrator will
prompt the user to set the external calibration device to the desired
value and press ENT when ready. The calibrator will then store its input
reading and step to the next calibration point.





The display shown above is instructing the user to set the external
calibration device to simulate a flow of 5,000 gallons per minute and
then press ENT.

If calibration procedure specifies approximate values, the calibrator will
prompt the user to set the external calibration device to the desired
value and then enter the exact value that was set. The calibrator will
then store its input reading and step to the next calibration point.





The display above is instructing the user to set the external calibration
device to simulate a flow of approximately 10,000 gallons per minute.
If you are able to set exactly 10,000 GPM, then you enter 10000 on the
Operation

230/235

17

mA Man_GPM
+4.000 5000
Set Ext Output
Press Ent

mA Man_GPM
+4.000 10000
Set Ext Output
Enter Value