If the alarm trip point is not within the desired accuracy, adjust
the alarm set point and repeat from step 2.
7.
Press the DEC key to ready the ramp at its ending (400°) value.
8.
Press the DEC key again to start the Ramp.
9.
Wait for the alarm to reset. The input display will again display
“tripped” and the Ramp will be stopped.
10. If the alarm release point is not within the desired accuracy,
adjust it and repeat from step 7.
Figure 5
Application Notes
230/235
65
NO NC
INPUT
- +