Martel Electronics 235 User Manual

Page 43

Advertising
background image

appear.


This menu allows the selection of engineering units for pressure
display. The choices are displayed 4 at a time. Scrolling past the first or
last displayed selection provides additional choices.

The engineering units available are: PSI, inH₂O, inHg, Kpa, cmH₂O,
mmHg, mbar, Bar. The selection between absolute and gauge is made
automatically based on the type of module attached.

When any of the water column scales are selected from the main
engineering units menu, a second menu will be displayed that allows
the operator to select the conversion temperature desired. 20°C is the
Instrument Society of America (ISA) standard, 60°F is used by the
American Gas Association (AGA) and 4°C is used in scientific research.
Software versions prior to 03.35 use the 4°C conversion temperature.

The factors used by the Pressure Module to convert from PSI to other
engineering units are listed below.

2.036

“Hg

@0°C

51.715

mmHg

@0°C

6.8948

KPascals

68.948

milliBars

0.068948

Bars

27.681

“H2O

@4°C

27.707

“H2O

@60°F

27.730

“H2O

@20°C

70.310

cmH2O

@4°C

70.376

cmH2O

@60°F

70.433

cmH2O

@20°C

When the OPT key is pressed a subsequent time(s), the calibrator
checks the ID and serial number of the attached module. If these
match the previous values, the module is activated immediately,
bypassing the title and serial number displays.
Input Functions

230/235

38

Engineering Unit
PSI “H2O “Hg Kpa

Advertising
This manual is related to the following products:

230