Ramp function – Martel Electronics 235 User Manual

Page 57

Advertising
background image

polarity of the current is significant and is the same as the voltage
terminals, i.e., red is positive. If the test current is of the incorrect
polarity, the calibrator will display REVERSED. Simply reverse the leads
at the calibrator or the devise under test.

Ramp Function

The ramp function provides an output ramp with adjustable rate and
amplitude for volts, millivolts, and thermocouples. Before a ramp can
be generated, select one of these output types. When the RAMP key is
pressed, the following screen will be seen:



Enter the starting value for the ramp in engineering units for the
function you are using and press ENT. The next screen will be:



Enter the ending value for the ramp in engineering units for the
function you are using and press ENT. The ending value must be
greater than the starting value.

The next screen will be:



Enter the time in seconds for the ramp to slew from the starting value
to the ending value. The range is 10 to 100,000 seconds. The next
screen will be:


If you want trip detect enabled for use with the ramp function,
highlight Enable and press ENT. If you do not want trip detect enabled,
highlight Disable and press ENT. See the description of Trip Detect for
more information on this function.
Output Functions

230/235

52

Starting value

??????

Ending value

??????

Ramp time
secs ??????

Trip Detect
Disable Enable