Pause (Set pause time for auto CalPts and Ramp)

The AUTO key will cause the Calibration Points or Ramp program to
repeat continuously without intervention. This setup option allows you
to specify how long the calibrator will pause between Calibration Points
or Ramps when the AUTO key is pressed.



Enter the dwell time in seconds to pause between each cal point or at
the end of each ramp cycle, and press ENT.

SerItf (Serial Interface) (Model 235 Only)

This enables the feature port on the bottom of the Model 235 to be
configured to calibrate the calibrator using the BetaCal Software or for
connection to the Calibration System.



Select BetaCal when calibrating the calibrator with BetaCal software.

For normal calibrator operation, select SYS. The selection of BetaCal or
SYS can be saved with the Save feature (see the following page).

MfgTst (Enable manufacturing test routines)
(Model 230 Only)

This option enables several test routines used in the manufacturing,
test, and calibration of the calibrator. These routines are not intended
for use in the field. A special factory access code is required to activate
these routines.



Operation

230/235

27

seconds btwn cal
points 10

Serial Intf Mode
BetaCal SYS