Calibration of mv to current converters – Martel Electronics 235 User Manual

Page 64

Advertising
background image

Application Notes

Calibration of mV to Current Converters

This example assumes a 10 to 50 mV input and a 4 to 20 mA output.

1.

Connect as shown in Figure 1.

2.

Select mA input.

3.

Select mV output.

4.

Enter 10 mV and adjust converter zero for 4 mA output.

5.

Enter 50 mV and adjust converter span for 20 mA output.

6.

Repeat steps 4 and 5 until both zero and span are within desired
tolerance.


















Figure 1

Application Notes

230/235

59








mV/Current

Converter

OUTPUT


- +





INPUT


- +

Input 10-50 mV

Output 4-20 mA

Advertising
This manual is related to the following products:

230