Martel Electronics 235 User Manual

Page 61

Advertising
background image

Enter the desired current in milliamps and press ENT. If an improper or
out of range value is entered, the calibrator will display an error
message. If the value is within the 0-22 mA operating limits, the
calibrator will set the output to the requested value. The output
displayed is the set point – not the actual output.

To generate a voltage output, press the V key while the calibrator is in
the output mode. (Note: the module should not be used for mV). The
calibrator will ask whether to use the 0-10 Volt output from the module
or the 0-1 volt output available at the calibrator’s output jacks. The
display will show:



Use the INC/DEC keys to select yes or no, and press ENT.

If you select the “yes” option to use the 10V output module, the
calibrator will accept values in the range of 0.0000 to 10.0000 and
output the requested values at the module volts output jacks.

If you select the “no” option to use the calibrator’s internal 0-1 volt
output, the calibrator will accept values in the range of 0.000000 to
0.100000 and 0.100001 to 1.00000 volts and output the requested
value at the calibrator’s volts output jacks. The module’s volts output
jacks will be activated.

Each time the volts option function is selected, the calibrator will give
the operator the opportunity to use the calibrator or the module
output provided that the module has been activated by pressing the
OPT key.

Each time that the milliamp or volts output function is selected, the
calibrator will check that the module is still attached and operating
properly. If the module does not respond properly, an error message
will be displayed.
Output Functions

230/235

56

Use 10V module?
yes no

Advertising
This manual is related to the following products:

230