Figure 10 – Martel Electronics 235 User Manual

Page 75

Advertising
background image

Transmitter Calibration Using Loop Power Supply (LPS)

























Figure 10









Application Notes

230/235

70











Pressure

Module

+

-

HI

LO

Vent to

Atmosphere

Special
mA Connection

Hand Pump

Instrument
Under Test

24VDC
4-20mA

Loop Power
Connection

Loop Power Supply

(LPS)

Stored in Pouch

Pressure Connection 1/8” FNPT