Figure 10 – Martel Electronics 235 User Manual
Page 75
Advertising

Transmitter Calibration Using Loop Power Supply (LPS)
Figure 10
Application Notes
230/235
70
Pressure
Module
+
-
HI
LO
Vent to
Atmosphere
Special
mA Connection
Hand Pump
Instrument
Under Test
24VDC
4-20mA
Loop Power
Connection
Loop Power Supply
(LPS)
Stored in Pouch
Pressure Connection 1/8” FNPT
Advertising
This manual is related to the following products: