2.3 Connecting Peripherals

2.3.1 Connection Terminals

Remove Back Cover:

Remove the screws from the four corners at the back of the conductivity
monitor. Remove the back cover. The connectors are exposed on the back
PCBA as shown in the Figure 1 below.

Connectors:

• J11 – 24V DC power connector
• J8 – 9V DC power connector (1.2mm diameter)
• J9 – Relay connections
• J10 - Conductivity electrode & Temperature probe connections (wiring has

to be done in the detachable connector

J10

J8

9V DC Power

Screw

Figure 1: Outer Side of Back PCBA

J10 Connections
1. CON Low Current
2. CON Low Voltage
3. CON High Voltage
4. CON High Current
5. Pt 100 Compensate
6. Pt 100 Sense
7. Pt 100 GND
8. (No

connection)


24V DC

J11

J11 Connections

1. +ve
2. Gnd

1)+ve
2)Gnd