Thermo Fisher Scientific Alpha COND 550 User Manual

Page 29

Advertising
background image

Instruction Manual

Alpha-

COND 550

23

measuring range is selected to ‘0.0 to 200.0 mS/cm’.

Press

▲ or ▼ key to set the line resistance of the sensor cable

connected to the monitor.

Press ENT key to confirm your selection.

4

Reset to factory defaults: The display shows ‘nO dEF’.

Press

▲ or ▼ key to select:

nO dEF : Not default reset of the monitor. User configured values

remain active, when confirmed by pressing ENT key.

FCt dEF:

Reset all settings of the monitor to factory defaults, when
confirmed by pressing ENT key.

CAL dEF:

Reset all calibration settings of the monitor to factory
defaults, when confirmed by pressing ENT key.

NOTE: Refer section 6 for factory/calibration default settings

Press ENT key to confirm your selection.

- If ‘nO dEF’ was selected, the monitor reverts to Cnfg screen.

Press

▲ or ▼ key to access other setup screens or press ▲

and

▼ key simultaneously (escape) to return to measurement

mode.

- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the monitor performs the
selected reset and returns to measurement mode.

Press

▲ or ▼ key to access other setup screens or press ▲ and ▼ key

simultaneously (escape) to return to measurement mode.

To exit from any intermediate steps, press

▲ and ▼ keys simultaneously

(escape). The monitor returns to the first screen configuration settings (Cnfg).