Instruction Manual

Alpha-

COND 550

22

5.5 Configuration Settings

Configuration settings let you select between 2-cell & 4-cell and reset the monitor to
factory defaults.

ENT

SETUP

HOLD

SETUP

HOLD

ENT

SETUP

HOLD

SETUP

HOLD

ENT

SETUP

HOLD

ENT

ENT

SETUP

HOLD

SETUP

HOLD

1

MEAS

°C

ATC

READY

µS

This screen appears only

when 2-cell is selected

and measuring range is set to

0.0 to 200.0 mS/cm

R.

2

2

3

4

4

4

ENT

ENT

1

1

From measurement mode press ENT key to enter setup mode as

described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press

▲ or ▼ key to select configuration settings screen

(Cnfg).

Press ENT key to access configuration settings (Cnfg).

2

Select cell type of the sensor: The lower display shows ‘CEL’. The

upper display shows the last configured cell type (‘2’ or ‘4’).

Press

▲ or ▼ key to select the cell type of your conductivity probe.

Press ENT key to confirm your selection.

NOTE: If you use a 2-cell conductivity sensor, make sure that the pin 1
and 2 are shorted and pin 3 and 4 are shorted on the J10 connector.
(Refer section 2.3.3 for more details)

3

Setting line resistance: The lower display shows ‘L.Ad’ and the upper

display shows the last configured line resistance; otherwise the default is
0.00 R.

NOTE: Line resistance of sensor cable constitutes to an error in
measurement of high conductivity values. The monitor will compensate
this error if you specify line resistance value in this setting. The setting is
available only if the cell type is set to 2-cell (step 3 above) and the

Advertising