5.6 Viewing Electrode Properties

Each time you calibrate the conductivity sensor, the monitor re-calculates the
calibration factor of the sensor and shown in the LCD at the end of calibration. The
setup mode allows you to view the last saved calibration factor and cell constant at
any time.

SETUP

HOLD

2

SETUP

ENT

1

ENT

HOLD

SETUP

HOLD

ENT

1

1

From measurement mode press ENT key to enter setup mode as

described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press

▲ or ▼ key to select viewing electrode properties screen

(CdAt).

Press ENT key to view electrode properties (CdAt).

2

Cell constant: The lower display shows ’CEL’ and the upper display

shows the last configured cell constant.

Press ENT key.

3

Calibration factor: The lower display shows ‘FCt’ and the upper display

shows the calibration factor based on the last conductivity calibration.

NOTE: The upper display shows ‘- - - ‘ if the monitor has not been
calibrated yet or the monitor has been reset to factory defaults.

Press ENT key to quit from electrode properties screen. The monitor
reverts to CdAt screen.

Press

▲ or ▼ key to access other setup screens or press ▲ and ▼ key

simultaneously (escape) to return to measurement mode.