4.2 Calibration

The monitor requires one-point calibration. Conductivity calibration is always
carried out in the specific measuring range selected for the monitor. Make
sure, the monitor is set to desired measuring range before you enter
calibration mode (Refer Section 5.4 for selecting measuring ranges).

Select a suitable standard conductivity solution so that its conductivity value
is within 10% to 100% of the full scale of the selected measuring range of the
monitor. (For better accuracy, select the calibration standard to be nearly
50% of the full scale of the measuring range selected.) Prepare calibration
standard solution in two beakers – one for rinsing and the other for calibration.
Prepare separate de-ionized water for electrode rinsing.

All new calibration values will automatically override existing calibration data.
If you wish to abort the calibration, press

▲ and ▼ keys simultaneously and

the monitor reverts to measurement mode and the monitor will revert to
previous calibration data.

CAL

HOLD

CAL

HOLD

MEAS

°C

ATC

READY

ENT

1

2

CAL

HOLD

3

ENT

HOLD

K=

2

CAL

HOLD

K=

µS

CAL

HOLD

4

READY

READY

OR

µS

CAL

HOLD

3

µS

READY

ENT

2

MEAS

°C

ATC

READY

µS

1

1

From measurement mode press CAL key to enter calibration mode.

2

The display briefly shows the cell type (“2 CELL” or “4 CELL”) to which

the monitor is configured (Refer section 5.5 for cell type settings)

Next, the last configured cell constant (k) is displayed in the upper
display. Use

▲ or ▼ keys to adjust the cell constant to mach the cell

constant of the conductivity probe. Adjustable range is 0.10 to 10.0.
Press ENT key
to confirm the new cell constant value.

Rinse the electrode in de-ionized water and then rinse with the
calibration standard. Immerse the probe in calibration standard solution
intended for calibration and swirl gently to create a homogenous sample.
Make sure there are no any air-bubbles trapped in the sensor.

3

The upper display shows the currently measured conductivity value of

the solution based on last calibration (if any). The lower display shows
uncalibrated (default) reading of the solution. The buffer annunciator
appears in LCD. The ‘READY’ annunciator appears when the reading is
stable.

Once the reading stabilizes press the

▲ or ▼ key to adjust the upper

display value to that of the standard solution. Press the ENT key to
accept the value.