2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors

The conductivity monitor supports both Pt100 & Pt1000 (2-wire or 3-wire)
temperature sensors. The default factory setting is Pt100. If you need to use
Pt1000 temperature sensor, you have to change the jumper setting (J7) as
described below.

Remove Back Cover:

Remove screws from the four corners at the back of the conductivity monitor.
Remove the back cover.

Remove Back PCBA:

Remove the screw located center of the back PCBA
(Figure 1). Detach the back PCBA from t
he monitor. Turn over the back PCBA.
Locate J7 jumper on the inner side of the back PCBA as shown in Figure 2
below.


Figure 2: Inner Side of Back PCBA

Set Jumper J7:

Set the J7 jumper to required sensor (Pt100 or Pt1000) type.