2.3.3 Connecting Conductivity Probe

Any industrial conductivity probe (2-cell or 4-cell) can be connected to the
conductivity monitor.

2-Cell Conductivity Probe:

1. Short Pin 1 & 2 of J10 connector using a jumper wire.

2. Connect CON low input wire (white) to Pin 2 of J10 connector.

3. Connect CON high input wire (black) to Pin 3 of J10 connector.

4. Short Pin 3 & 4 of J10 connector using a jumper wire.

4-Cell Conductivity Probe:

1. Connect CON low input wire (black) to Pin 1 of J10 connector

2. Connect CON low input wire (white) to Pin 2 of J10 connector.

3. Connect CON high input wire (orange) to Pin 3 of J10 connector.

4. Connect CON high input wire (blue) to Pin 4 of J10 connector.

2.3.4 Connecting Temperature Probe

For Automatic Temperature Compensated (ATC) conductivity readings, an
in-built temperature sensor is usually integrated with industrial conductivity
probes. Alternatively, a separate 100

Ω Pt RTD temperature probe (2-wire or

3-wire) can be connected to the monitor.

3-Wire Probe:

1. Connect PT100 compensate wire to Pin 5 of J10 connector.

2. Connect PT100 sense wire to Pin 6 of J10 connector.

3. Connect PT100 GND wire to Pin 7 of J10 connector.

2-Wire Probe:

1. Short Pin 5 & 6 of J10 connector using a jumper wire.

2. Connect PT100 sense wire to Pin 6 of J10 connector

3. Connect PT100 GND wire to Pin 7 of J10 connector.