Enter k factor unit, Enter meter's k factor, Select rate/total units of measure – Badger Meter B2800 User Manual

Page 11: Select display function, Totalizer pulse output, Password

Advertising
background image

At the METER prompt, press UP or RIGHT arrows to select the bore size of your meter Press ENTER to save meter size choice

and advance to the K factor units selection

OTEE:

N

The meter connection size and the bore size are different For example, many of the 1 inch NPT turbines have bore

sizes that range from 3⁄8 inch up to 1 inch Be sure to use the correct bore size or the meter could report incorrect

flows and totals

Enter K Factor Unit

Directly after the METER size is selected, the display's K factor unit must be chosen Use the UP arrow to select the K factor

unit For meters calibrated in gallons, use PUL/GAL (pulses per gallon); for meters calibrated in cubic meters, use PUL/M3

(pulses per cubic meter) and so forth Press ENTER to save the K factor unit and advance to the next parameter

OTEE:

N

Unless otherwise specified, turbine flow meters are supplied with K factors measured in pulses per gallon (PUL/GAL),

which will automatically convert to your desired units of measure

OTEE:

N

The K factor supplied with the meter or calculated from calibration data will be needed to complete the next step

Enter Meter's K Factor

To change the K FACTOR value, use the RIGHT arrow and select the position of the number to be changed Using the UP

arrow, increment the display digit until it matches the meter’s K factor digit Repeat this process until all K factor digits have

been entered Press ENTER to save the K factor and advance to the RATE/TOTAL unit selection

Select Rate/Total Units of Measure

The monitor allows the choice of five common rate/total units The monitor indicates the rate/total unit of measure the

display is currently set for If the current selection is correct, press ENTER to advance to the next parameter To change to an

alternate unit, use the arrow keys to scroll to the desired rate unit and press ENTER to save the choice

Selection

Rate

Total

GPM / GAL

Gallons per minute

Gallons

LPM / LIT

Liters per minute

Liters

M3PH / M3

Cubic meters per hour Cubic meters

M3PD / M3

Cubic meters per day

Cubic meters

BPD / BBL

Oil barrels per day

Oil barrels

OTEE:

N

The total unit’s output multiplier cannot be modified in the simplified program level

Select Display Function

The monitor can display RATE or TOTAL or alternate between BOTH rate and total If the current selection is correct, press

ENTER to advance to the next parameter To change to an alternate display mode, use the arrow keys to scroll to the desired

display mode and press ENTER to save the choice
A TEST function is also available in the display function submenu With the test function selected, the display acts like a

frequency counter and displays the raw input frequency being supplied to the frequency input terminals This is very useful

when troubleshooting flow problems

Totalizer Pulse Output

The pulse output parameter can be either enabled or disabled When enabled this output generates 20 mS duration pulse

every time the least significant digit of the totalizer increments (20 Hz maximum) The amplitude of the pulse is dependent on

the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC

Password

Password protection prevents unauthorized users from changing programming information Initially, the password is set

to all zeros To change the password, simply enter any four digit code using the arrows as previously described to enter the

password value Pressing ENTER will store the password and take you back to the RST PSWD screen

OTEE:

N

This password will allow users to reset totals

Programming Manual