Turbine Flow Sensor

Installation & Operating Instructions

Page 10

03/12 Form No. 05-TUR-UM-00194

DO - Choose a position for the flow sensor that will not be at the lowest level in the system. Placing the

flow sensor at a higher elevation in the system will avoid collection of debris, sediment and dirt in the flow

sensor.

DO - Use a filter. All applications should be filtered to at least 40 micron.

DON’T - Locate a flow sensor directly in-line with the outlet of a pump, as pressure pulsations can react

with the turbine. Locate the sensor after another component, observing the 10 port diameter rule.

DON’T - Adjust the magnetic pick-up on the flow sensor. This is calibrated at the factory. Further adjustment

will cause a decrease in performance or damage to the sensor.

DON’T - Exceed the working temperature range of –4 °F to +300 °F (–20 °C to +150 °C). Higher

temperatures will damage the magnetic pick-up and lower temperatures will limit the rotation of the

turbine.

Electrical Connections - Standard Magnetic Pick-up

(BLACK)

-

(WHITE) N.C.

(RED)

+

Cable Assembly

F6234-6

6 Ft.

F6234-15 15 Ft.

Top of

F6234 Cable

3

2

1

1 - (WHITE) N.C.

2 - (BLACK) -

3 - (RED) +

Magnetic Pick-up

Male Connector

1

2

3

(BLACK)

-

(WHITE) N.C.

(RED)

+

Cable Assembly

F2832-6

6 Ft.

F2832-15 15 Ft.

A -

(RED)

+

B -

(BLACK)

-

Top of

F2832 Cable

B A

Magnetic Pick-up

Male Connector

B

A

FIGURE 2

Standard Magnetic Pick-up with Frequency Output – 2-pin Connector

FIGURE 3

Standard Magnetic Pick-up with Frequency Output – 3-pin Connector