Electrical Connections - IFC (Intelligent Frequency Converter)

The 4-20 mA output can drive auxiliary devices (resistive

loads) such as displays, recorders and computers, provided

that the voltage supplied by the power supply is adequate.

Devices must be wired in series with the F to I converter

and power supply. The voltage drop across the load(s) and

the 6 Vdc minimum needed to drive the F to I converter

determine the minimum voltage required from the power

supply.
Determine the necessary voltage required to adequately

drive the F to I converter and auxiliary device(s).
The F to I converter acts as a current controlling device.

Thus, the current output remains the same even if the power

supply voltage fluctuates or the load resistance changes.

The current varies only with respect to the flow rate from

the turbine flow sensor, as long as the voltage drop across

the F to I converter is at least 6 Vdc.
The load(s) in the circuit will generally have some electrical

resistance, 100 Ohms for this example. The 4-20 mA loop

current will produce a voltage drop across each load. The

maximum voltage drop across a load(s) will exist when

the loop current is 20 mA. The power supply must provide

enough voltage for the load(s) plus the 6 Vdc minimum

insertion loss of the F to I converter.

F to I

Converter

24 VDC

Power

Supply

4 - 20 mA

150

Ohms

100

Ohms

50

Ohms

Example 1

Sufficient Power Supply Voltage

Total Load Resistance = 300 Ohms

At 20 mA loop current, the voltage drop across the load(s)

is 6 volts:

300 Ohms × 20 mA = 6,000 mV or 6 volts

Subtract 6 volts from the 24 volt source to determine that

18 volts is available to power the F to I converter. The 18

volts is within the specified 10 to 30 volt range and is

sufficient to power the F to I converter.

F to I

Converter

24 VDC

Power

Supply

4 - 20 mA

1000

Ohms

Example 2

Insufficient Power Supply Voltage

Total Load Resistance = 1000 Ohms

At 20 mA loop current, the voltage drop across the load(s)

is 14 volts:

1000 Ohms × 20 mA = 20,000 mV or 20 volts

Subtract 20 volts from the 24 volt source to determine that

4 volts is available to power the F to I converter. The 4

volts is below the specified 10 to 30 volt range and is not

adequate to power the F to I converter. If for example, the

power supply voltage was 30 volts instead of 24 volts, the

voltage available to power the F to I converter would be 10

volts and within the specified range.

Cable

Connector

3

4

5

2

1

F to I Converter

Male Connector

3

2

1

4

5

+ 4 - 20 mA

- 4 - 20 mA

+ 4 - 20 mA (Sink)
- 4 - 20 mA (Source)

N.C.
N.C.
N.C.

PIN 3
PIN 4
PIN 5

PIN 2

PIN 1

No Connection

RED
BLACK
WHITE

(RED)

+ Loop

(WHITE) N.C.

(BLACK)

- Loop

Cable Assembly

F6557-6

6 Ft.

F6557-15 15 Ft.

FIGURE 4

IFC with 4-20 mA Output (F to I) – 5-pin Connector