Datalogger, Datalogger 16, Datalogger store delete – Badger Meter MC4000 Diagnostic Test Equipment User Manual

Page 16: Datalogger store on, Datalogger store off

Advertising
background image

oPerATIon

P1

bar

74.32

T

°C

86.5

Q

l/min

38.4

P2

bar

76.12

Datalogger

Recording Measurements

The internal memory can save all displayed measurements The USB data port permits the uploading of the stored data to a

PC where for further processing The internal memory permits 224 individual records with a total capacity of 2 5 MB Use the

MC4000 keyboard to initializes each individual record After initializing the record location, pressing ESC switches the display

into measuring mode In measuring mode, an M icon indicates an active datalogging session

P1

bar

74.32

T

°C

86.5

Q

l/min

38.4

P2

bar

76.12

M

Logging Active

Indication

Datalogging Memory Commands

The first menu screen allows a choice to start, stop or delete the datalogger memory

DATALOGGER

STORE DELETE

This command erases datalogger memory

All records are deleted

DATALOGGER

STORE ON

This command starts the recording of data at the selected

interval rate Each record automatically adds the date and time

from an internal real time clock

DATALOGGER

STORE OFF

Stops recording the datalogging information to memory

Page 16