MC4Con SoFTWAre

Pressure

Pressure parameter settings.

Scale P1

Full scale for the first pressure sensor (P1)

Scale P2

Full scale for the optional pressure sensor (P2)

Save Interval

Sampling and recording rate during data collection

Threshold

The threshold signal level for the P1 (Transient) Defines the point at which transient

recording starts

RPM

RPM parameter settings.

Scale rpm

Multiplication constant for rpm scaling

DScale rpm

Division constant for rpm scaling

New-Tab

Three point linearizing of flow sensors.

example

Tab F1 INP

First frequency point

0 Hz

Tab Q1 OUT

First displayed flow reading

0 LPM

Tab F2 INP

Second frequency point

166 Hz

Tab Q2 OUT

Second displayed flow reading

38 LPM

Tab F3 INP

Third frequency point

630 Hz

Tab Q3 OUT

Third displayed flow reading

150 LPM

Units

Flow units are calculated in US or European standard units.

EU

Default units are lpm, bar, ° C

US

Default units are gpm, psi, ° F

Language

Selection of English or default (the local country language).

English
Default

Backlight

Backlight control.

Light off

Backlight is turned off No instrument backlight

Light on

Backlight is turned on The display illuminates for 30 seconds when any buttons are pressed

File

Menu MC4000, upper tree switches.

New Device

Default setting

File Read

Reads all the menu parameters from a file with the file format filename set from the MC4000

File Save

Stores all the menu parameters in a file with the format filename set to the MC4000

Device

Menu MC4000, lower three switches.

Menu MC4000 read

Reads the stored MC4000 parameters into the MC4CON software

Menu MC4000 store

Reads the stored MC4CON parameters into the MC4000

Timer MC4000 write

Transfer time and date values from the PC into MC4000

Page 26