SUPPleMenTArY MeASUreMenTS

Three Point Flow Sensors Linearizing Using the New Tab Function

Linearization achieves better accuracy of the flow sensor connected to the MC4000 The MC4000 is capable of using up to

three linearization points The points entered are in pairs of frequency values with their associated flow rates The linear point

pairs can be entered using the MC4000 keyboard or from a connected PC

Linearization Point

Frequency (Hz)

Flow Rate (Q)

1

Frequency F1

Flow Rate Q1

2

Frequency F2

Flow Rate Q2

3

Frequency F3

Flow Rate Q2

Table 1: Sample linearization table

The linearization menu makes the assignment of both the frequencies and display readings

example

Linearization Pair Number

Frequency Values

Flow Values

Linearization Pair 1

0 Hz

0 lpm

Linearization Pair 2

166 Hz

66 lpm

Linearization Pair 3

630 Hz

150 lpm

Table 2: Sample linearization values