Badger Meter MC4000 Diagnostic Test Equipment User Manual
Page 21
SUPPleMenTArY MeASUreMenTS
Three Point Flow Sensors Linearizing Using the New Tab Function
Linearization achieves better accuracy of the flow sensor connected to the MC4000 The MC4000 is capable of using up to
three linearization points The points entered are in pairs of frequency values with their associated flow rates The linear point
pairs can be entered using the MC4000 keyboard or from a connected PC
Linearization Point
Frequency (Hz)
Flow Rate (Q)
1
Frequency F1
Flow Rate Q1
2
Frequency F2
Flow Rate Q2
3
Frequency F3
Flow Rate Q2
Table 1: Sample linearization table
The linearization menu makes the assignment of both the frequencies and display readings
example
Linearization Pair Number
Frequency Values
Flow Values
Linearization Pair 1
0 Hz
0 lpm
Linearization Pair 2
166 Hz
66 lpm
Linearization Pair 3
630 Hz
150 lpm
Table 2: Sample linearization values
Page 21
Month 2014