Register Mappings

Data Component Name

MODBUS Registers

Available Units

Long Integer

Format

Single Precision

Floating Point Format

Spare

40100…40101

40200…40201

Flow Rate

40102…40103

40202…40203

Gallons, Liters, MGallons, Cubic Feet,

Cubic Meters, Acre Feet, Oil Barrel, Liquid

Barrel, Feet, Meters, Lb, Kg, BTU, MBTU,

MMBTU, TON

Per

Second, Minute, Hour, Day

Spare

40104…40105

40204…40205

Positive Totalizer

40106…40107

40206…40207

Grand Total Totalizer

40108…40109

40208…40209

Battery Voltage

40110…40111

40210…40211

x.xx

Spare

40112…40113

40212…40213

Table 5: Modbus register map for "Little-Endian" word order master devices

For reference: If the ER-500 Totalizer = 12345678 hex

Register 40106 would contain 5678 hex (Word Low)

Register 40107 would contain 1234 hex (Word High)

Data Component

Name

MODBUS Registers

Available Units

Long Integer

Format

Single Precision

Floating Point Format

Spare

40600…40601

40700…40701

Flow Rate

40602…40603

40702…40703

Gallons, Liters, MGallons, Cubic Feet,

Cubic Meters, Acre Feet, Oil Barrel, Liquid

Barrel, Feet, Meters, Lb, Kg, BTU, MBTU,

MMBTU, TON

Per

Second, Minute, Hour, Day

Spare

40604…40605

40704…40705

Positive Totalizer

40606…40607

40706…40707

Grand Total Totalizer

40608…40609

40708…40709

Battery Voltage

40610…40611

40710…40711

x xx

Spare

40612…40613

40712…40713

Table 6: Modbus register map for "Big-Endian" word order master devices

For reference: If the ER-500 Totalizer = 12345678 hex

Register 40606 would contain 1234 hex (Word High)

Register 40607 would contain 5678 hex (Word Low)

ER-500 Flow Monitor

Page 34