Totalizer Reset

The Job Totalizer can be reset by momentarily contacting the Total Reset terminal to ground or

pressing the MENU and ENTER buttons simultaneously The Grand Totalizer can be reset via selection

in the Advanced Menu or Modbus

Totalizer Preset

User can preset Job Total values

Safety

Certifications

Class I Division 1, Groups C, D; Class II, Division 1 Groups E, F, G; Class III for US and Canada Complies

with UL 913 and CSA C22 2 No 157-92

Entity

Parameters

4-20mA

Loop

V

max

= 28V DC I

max

= 26 mA

C

i

= 0 5 μF

L

i

= 0 mH

Pulse

Output

V

max

= 28V DC I

max

= 100 mA C

i

= 0 μF

L

i

= 0 mH

Reset Input

V

max

= 5V DC

I

max

= 5 mA

C

i

= 0 μF

L

i

= 0 mH

RS485

V

max

= 10V DC I

max

= 60 mA

C

i

= 0 μF

L

i

= 0 mH

Sensor

Input

V

oc

= 2 5 V

I

sc

= 1 8 mA

C = 1 5 μF

L

a

= 1 65 H

EMC

2004/108/EC

Accuracy

0 05%

Response Time

1…100 seconds response to a step change input, user adjustable

Environmental

Limits

–22…158° F (–30…70° C); 0…90% humidity, non-condensing

Materials

Polycarbonate, stainless steel, polyurethane, thermoplastic elastomer, acrylic

Enclosure

Ratings

NEMA 4X/IP 66

Engineering

Units

Gallons, Liters, Oil Barrels (42 gallon), Liquid Barrels (31 5 gallon), Cubic Meters,

Million Gallons, Cubic Feet, Million Liters, Acre Feet

Rate Time

Seconds, minutes, hours, days

Totalizer

Exponents

0 00, 0 0, X1, x10, x100, x1000

K factor Units

Pulses/Gallon, Pulse/cubic meter, pulses/liter, pulses/cubic foot

ER-500 Flow Monitor