4 Basic Operations

3

Remove the objective lens cap and turn the retinal
camera's power switch to the ON position.

The POWER lamp lights and the set lamp blinks.

The CR-2 Plus AF automatically does an initialization sequence

(internal check and initialization of driving parts). After the

initialization the parts of the main unit stop in their center

positions.

CAUTION

The main unit and the chin rest move to the center position when the CR-2 Plus AF is turned on.

Do not touch the main unit or the chin rest while they are moving. Keep the patient’s chin away

from the main unit and chin rest when they are moving.

Power-saving function

If the power is on and the buttons on the operation panel are not operated for over 10 minutes, the

power-saving function is activated. The POWER lamp blinks while in power saving mode.

To return to normal mode, press any button on the operation panel or move the stage right or left.

4

Press and hold the unlock button and slide the stage
lock to the UNLOCK side.

Align the mark on the stage lock and the mark on the left side

of the base.

The stage lock is released.

5

Move the stage towards you with the operation lever.

6

Disinfect the forehead rest and replace the chin rest paper.

If chin rest paper is not being used, disinfect the chin rest also.

CAUTION

t To prevent the risk of infection, wipe the forehead rest with an ethanol solution for each patient.

For details on how to disinfect, consult a specialist.

t To ensure cleanliness, replace the chin rest paper for each patient.