4 Basic Operations

27

When the top and the bottom of the image of the pupil are unified, the anterior segment automatically

switches to the fundus on the observation monitor.

ON

ON

Auto-fundus Observation Indicator

Displayed when Auto-fundus observation function is available.

Icon colors indicate the status of the detecting pupil.

Total color of icon is dark gray :

Error in the detecting pupil.

Indication bar is totally dark gray :

The detecting pupil is impossible.

The left side of indication bar is green : The position of the detecting pupil is not effective.

Indication bar is totally green :

The position of the detecting pupil is effective.

When the pupil is smaller than the inner pupil alignment circle, images cannot be taken

properly.

Darken the room further to dilate the patient's pupil more. If the patient's pupil does not dilate

enough, use the small pupil photography function (see page 33).

Adjustment of observation light brightness

Adjust the observation light brightness with the observation

light brightness button.

Guidelines for observation light brightness

Adjust to the lowest possible brightness level to reduce the

strain on the patient’s eye.

Observation monitor brightness adjustment

Use the Rics NM 2 to adjust the brightness of the observation monitor. See the Rics NM 2 manual

for details.