Canon CR-2 PLUS AF User Manual

Page 48

Advertising
background image

48

Appendix

Guidance and Manufacturer’s Declaration for Electromagnetic Immunity

The CR-2 Plus AF is intended for use in the electromagnetic environment specified below.

The customer or the user of the CR-2 Plus AF should assure that it is used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level

Electromagnetic Environment – Guidance

Electrostatic discharge

(ESD)

EN IEC 61000-4-2

±(2, 4, 6) kV contact

±(2, 4, 8) kV air

±(2, 4, 6) kV contact

±(2, 4, 8) kV air

Floors should be wood, concrete or ceramic

tile.

If floors are covered with synthetic material,

the relative humidity should be at least 30%.

Electrical fast transient/

burst

EN IEC 61000-4-4

±2 kV for power supply

lines

±1 kV for input/output

lines

±2 kV for power supply

lines

±1 kV for input/output

lines

Mains power quality should be that of a

typical commercial or hospital environment.

Surge

EN IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be that of a

typical commercial or hospital environment.

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines

EN IEC 61000-4-11

<5% U

T

(>95% dip in

U

T

) for 0.5 cycles

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 sec

<5% U

T

(>95% dip in

U

T

) for 0.5 cycles

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 sec

Mains power quality should be that of a

typical commercial or hospital environment.

If the user of the CR-2 Plus AF requires

continued operation during power mains

interruptions, it is recommended that

the CR-2 Plus AF be powered from an

uninterruptible power supply.

Power frequency

(50/60Hz) magnetic

field

EN IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be

at levels characteristic of a typical location in

a typical commercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Advertising