Page 103: Testing the analogue outputs

Advertising
background image

HMP230 SERIES

M210225en-B

Appendix 1: Serial commands

97

4.

OTHERS

ITEST

Testing the analogue outputs

ITEST <cr>

or

ITEST a b <cr>

a

=

current/voltage of channel 1

b

=

current/voltage of channel 2

The operation of the analogue outputs can be tested by forcing the outputs to
given values which can then be measured with a current/voltage meter from
the analogue outputs. The response to ITEST command gives six out-
puts/parameters. Only the first two are relevant; they show the channel current
or voltage in mA or V. The other four figures contain information for service
purposes only.

Examples:

reading the channel outputs and parameters

>itest <cr>

1.9438

2.3483

1.00694 10.64634

1.97374

2.17665

>

forcing outputs 0.5 V and 4 V to channels 1 and 2

>itest 0.5 4 <cr>

0.5000

4.0000

1.00694 10.62970

1.23336

3.01722

>

releasing the forced control and reading the outputs

>itest <cr>

1.9427

2.3392

1.00731 10.62428

1.97157

2.16978

>