HMP230 SERIES
User's Guide

M210225en-B

58

•

Give command FCRH <cr>, enter the first point value and press <cr>:

>FCRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...

•

The stabilization of the sensor to the reference humidity can be
monitored by giving c <cr>:

RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

•

Insert the sensor head into a measurement hole of the NaCl chamber.

•

Wait at least 30 minutes.

•

Press any key and enter the second point value and press <cr>.

RH : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

7.1.3.2. With display/keypad commands

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Disconnect the security lock jumper!

•

Insert the sensor head into a measurement hole of the LiCl chamber.

•

Wait at least 30 minutes.

•

Select Cali in the main menu and then RH cal; select Sensor
changed. Change the first point reading with the arrow keys and
press ENT.

•

Insert the sensor head into a measurement hole of the NaCl chamber.

•

Wait at least 30 minutes.

•

If necessary, change the second point reading with the arrow keys and
press ENT.

7.1.3.3. With LED commands

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Disconnect the security lock jumper!

•

Insert the sensor head into a measurement hole of the LiCl chamber.