Analogue output commands, Amode setting the analogue outputs – Vaisala HMP230 User Manual

Page 87

Advertising
background image

HMP230 SERIES

M210225en-B

Appendix 1: Serial commands

81

1.

ANALOGUE OUTPUT COMMANDS

AMODE Setting the analogue outputs

Disconnect the security lock jumper!

AMODE a bb.bbb cc.ccc d ee.eee ff.fff <cr>

a

=

channel 1:

U =

voltage output

I

=

current output

bb.bbb =

lower limit of channel 1

cc.ccc =

upper limit of channel 1

d

=

channel 2:

U =

voltage output

I

=

current output

ee.eee =

lower limit of channel 2

ff.fff

=

upper limit of channel 2

The bb.bbb, cc.ccc, ee.eee and ff.fff parameters are entered in volts or milli-
amperes.

Sets the analogue outputs on channels 1 and 2. An example: the voltage output
on channel 1 is set to be 0...1 V and channel 2 set to 2...10 V:

>AMODE U 0 1 U 2 10 <cr>
Ch1 :

0.000 ...

1.000 V

Ch2 :

2.000 ...

10.000 V

The current settings can be checked by giving the command without any pa-
rameters:

>AMODE <cr>
Ch1 :

0.000 ...

20.000 mA

Ch2 :

0.000 ...

20.000 mA