HMP230 SERIES
Appendix 1: Serial commands

M210225en-B

84

2.

CALIBRATION COMMANDS

CRH

Relative humidity calibration

Disconnect the security lock jumper!

CRH <cr>

With this command the transmitters can be calibrated against a reference. A
one-point calibration can be done against an accurate transfer standard in the
field and a two-point calibration using saturated salt solutions in controlled
conditions. A two-point calibration is performed as follows:

>CRH <cr>
RH

:

12.00

Ref1 ?

11.3 <cr>

Press any key when ready ...
RH

:

76.00

Ref2 ?

75.5 <cr>

In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:

>CRH <cr>
RH

:

12.80

Ref1 ?

11.3 <cr>

Press any key when ready ...
RH

:

75.50

Ref2 ?

<cr>

If the stabilization of the sensor to the humidity in the calibrator needs to be
monitored, the measurement output can be repeated by giving command c<cr>
at Ref1 and Ref2:

>CRH <cr>
RH

:

12.00

Ref1 ?

c <cr>

RH

:

11.70

Ref1 ?

c <cr>

RH

:

11.50

Ref1 ?

11.3 <cr>

Press any key when ready ...
RH

:

76.00

Ref2 ?

75.5 <cr>

FCRH

Relative humidity calibration after sensor change

Disconnect the security lock jumper!

FCRH <cr>

After humidity sensor change the transmitter must be calibrated using this
command and the calibration must be done at two reference points. The cali-
bration is performed as follows:

>FCRH <cr>
RH

:

1.90

Ref1 ?

11.3 <cr>

Press any key when ready ...
RH

:

76.30

Ref2 ?

74.9 <cr>