If there are no error messages, only a prompt is displayed:

>ERRS <cr>
>

If errors have occurred, the transmitter outputs the error code (see Appendix 5
for error messages):

>ERRS <cr>
E40

f

( all )

out of range

>

ECHO

Turning the serial interface echo ON/OFF

ECHO xxx <cr>

xxx

=

ON or OFF

When the echo is off, the commands given through the serial interface or the
prompt > cannot be seen on the display.

When the serial interface is in half-duplex mode, the echo is always off. Even
then the ECHO command can indicate that echo is on.

INTV

Setting the output interval for the RUN state

INTV xxx yyy <cr>

xxx

=

output interval (0...255)
0: no pause between outputs

yyy

=

unit (s, min or h)

Sets the output interval when the transmitter outputs measurement readings to
a peripheral device.

For example, the currently valid settings are output with:

>INTV <cr>
Output intrv. :

0 min

When this is changed into 10 minutes, the command is:

>INTV 10 <cr>
Output intrv. :

10 min

The unit is changed into seconds with:

>INTV S <cr>
Output intrv. :

10 s

The change can also be done with one command:

>INTV 10 S <cr>
Output intrv. :

10 s