Vaisala HMP230 User Manual

Page 68

Advertising
background image

HMP230 SERIES
User's Guide

M210225en-B

62

7.2.2. Two point temperature calibration

7.2.2.1. With serial commands

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the lower reference.

•

Give command CT <cr>, enter the first point value and press <cr>:

>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready

•

If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:

T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...

•

Change the temperature and check the transmitter against the higher
reference.

•

Check that the reading corresponds with the reading of the reference
instrument. If not, adjust the second point

•

Press any key, enter the second point value and press <cr>.

T : xx.x Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

7.2.2.2. With display/keypad commands

•

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Disconnect the security lock jumper!

•

Check the transmitter against the lower reference.

•

Select Cali in the main menu and then T cal; select two-point
calibration T 2 point cal. Change the first point reading with the
arrow keys and press ENT.

•

Change the temperature and check the transmitter against the higher
reference.