Figure 4-3. numerical printout – MagTek InSpec 9000 User Manual

Page 43

Advertising
background image

Section 4. Graphics Presentations

33


InSpec 9000 numerical test results

Description:

Date and Time Printed: 1/29/98 10:41:18 AM

Track No.:

1

2

3


Min

Avg

Max

Min

Avg

Max

Min

Avg

Max


Amplitude%: 125 135

141

105 111 117

109 113 117


Bit Size%:

-11.7 -0.4 12.7

-10.0 -0.6 11.6

-10.9 -0.3 13.5


Sub

Interval%:

3.2 3.0 1.1

-3.3 -0.7 2.0

4.2 -0.4 2.7


Adj. Bit Size%:

-11.8 0.0 14.0

-11.0 0.0 12.0

-11.6 0.0 12.7


Adj. Sub Bit Size%:-3.8

0.0

3.9

-3.0 0.0 2.7

-4.8 0.0 3.9


Start Sentinel (in):

0.301 in

0.303

in

0.301

in


Start Sentinel (mm)

7.652 mm

7.702 mm

7.645 mm


Green - Within ISO Limits for New Card.
Yellow - Within ISO Limits for Returned Card.
Red - Exceeds All ISO Limits

Limits

Track No.:

1

2

3

Red

Green Red Red

Green Red Red

Green

Red

Yellow Yellow Yellow

Yellow Yellow Yellow


Amplitude%:

52 64 126 136

52 64 126 136

52

64 126 136


Bit Size%:

-15 -10 10 15

-15 -7

7 15 -15 -10 10 15


Sub Interval%:

-20 -12 12

20

-20 -10 10 20 20 -12 12 20


Adj. Bit Size%: -15 -10 10 15

-15 -10 10 15

-15 -10 10 15


Adj. Sub Bit Size%: -30 -12 12 30

-30 -12 12 30

-30 -12 12 30


Start Sentinel (in): 0.123 0.273 0.313 0.493 0.223 0.273 0.313 0.493 0.123 0.253 0.333 0.493

Start Sentinel (mm): 3.12 6.93 7.95 0.693 5.66 6.93 7.95 12.52

3.12 6.43 8.46 12.52


Measurement uncertainty of the InSpec 9000 System: Average amplitude +/-15%; Average bit size +/-2%; Start Sentinel +/-.007 inches.
To maintain the range of measurement uncertainty, the tester and card must be clean.
The card must also be flat and within ISO size specifications.
This product tests Amplitude, Jitter, Start Sentinel, and Baseline Integrity. There are other parameters that may influence read

reliability.

Before testing for limits, the program removes 0.130 inch from each end of the data.

Figure 4-3. Numerical Printout