35


To obtain an illustration of the Start Sentinel location, click on the “Start Sent.” button on the
ISO Parameters screen. The location and the numerical data will be shown on the screen as
illustrated in Figure 4-5, but the screen cannot be printed.

Figure 4-5. Start Sentinel


The ISO range of the start sentinel for Tracks 1 and 2 is 0.293

± 0.020 inch. For Track 3 the

range is 0.293

± 0.040 inch. The vertical line through the bar illustrates where the sentinel starts

in relation to the ISO standard. If the vertical line is within the green range, it is within spec.
The yellow range indicates that the start sentinel is out of range but will still be readable in most
applications.