Aalborg DFM Digital User Manual

Page 11

Advertising
background image

7

4. SPECIFICATIONS

FLOW MEDIUM: Please note that DFM Mass Flow Meters are designed to work

with clean gases only. Never try to measure flow rates of
liquids with any DFM.

CALIBRATIONS: Performed at standard conditions [14.7 psia (101.4 kPa) and

70

F

F (21.1

F

C)] unless otherwise requested or stated.

ENVIRONMENTAL (PER IEC 664):

Installation Level II; Pollution Degree II.

FLOW ACCURACY FOR DFM 27/37/47 (INCLUDING LINEARITY):

0

F

C to 50

F

C and 5 to 100 psia (34.5 - 689.5 kPa): ±1% of

full scale (F.S.)

FLOW ACCURACY FOR DFM 26/36/46 (INCLUDING LINEARITY):

±1% of FS at calibration temperature and pressure.

REPEATABILITY: ±0.15% of full scale.

FLOW TEMPERATURE COEFFICIENT:

0.15% of full scale/

F

C or better.

FLOW PRESSURE COEFFICIENT:

0.01% of full scale/psi (6.895 kPa) or better.

FLOW RESPONSE TIME:

DFM 26/27: 300ms time constant; approximately 1 second
to within ±2% of set flow rate for 25% to 100% of full
scale flow.

DFM 36/46/37/47: 600ms time constant; approximately 2
seconds to within ±2% of set flow rate for 25% to 100% of
full scale flow.

MAXIMUM BURST PRESSURE:

DFM 26/36/46: 1000 psig (6895 kPa gauge).
DFM 27/37/47: 200 psig (1379 kPa gauge).

PRESSURE MEASUREMENT RANGE:

0 to 100 psia (689.5 kPa absolute).
P (absolute) = P (gauge) + P (atmospheric)