Aalborg DFM Digital User Manual

Page 9

Advertising
background image

5

PIN FUNCTION

1

+15 Vdc (Optional +12 or +24 Vdc) Power Supply

2

0 5 Vdc or 4-20mA Flow Signal Output

3

0 5 Vdc or 4-20mA Temperature Signal Output (Optional)

4

0 5 Vdc or 4-20mA Pressure Signal Output (Optional)

5 (reserved)
6 (reserved)
7 (reserved)
8

Relay No. 1 - Common Contact

9

Relay No. 1 - Normally Open Contact

10

Relay No. 2 - Normally Closed Contact

11

RS485 (-) (Optional RS232 TX)

12 (No

Connection)

13 Common
14

-15 VDC Power Supply (Only for ±15Vdc option)

15

Common, Signal Ground For Pin 2 (4-20 mA return)

16

Common, Signal Ground For Pin 3 (4-20 mA return)

17

Common, Signal Ground For Pin 4 (4-20 mA return)

18

Common, Power Supply (- DC power for 12 and 24 Vdc)

19 Common
20

RS232 Signal GND (RS-485 GND Optional)

21

Relay No. 1 - Normally Closed Contact

22

Relay No. 2 - Common Contact

23

Relay No. 2 - Normally Open Contact

24

RS485 (+) (Optional RS232 RX)

25

Chassis Ground

Figure b-1, DFM 25 Pin "D" Connector Configuration.

IMPORTANT NOTES: In general, "D" Connector numbering patterns are
standardized. There are, however, some connectors with nonconforming
patterns and the numbering sequence on your mating connector may
or may not coincide with the numbering sequence shown in our pin
configuration table above. It is imperative that you match the appropriate
wires in accordance with the correct sequence regardless of the
particular numbers displayed on the mating connector.

Make sure power is OFF when connecting or disconnecting any cables
in the system.

The (+) and (-) power inputs are each protected by a 400mA M (medium time-lag)
resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut
power to the flow transducer circuit. Disconnect the power to the unit, remove the
faulty condition, and reconnect the power. The fuse will reset once the faulty con-
dition has been removed. DC Power cable length may not exceed 9.5 feet (3
meters).