Aalborg DFM Digital User Manual

Page 38

Advertising
background image

34

GAS FLOW

26 FlowOutScaleV

- DAC 0-5/0-10 Analog Output Scale for Flow

27 FlowOutScale_mA - DAC 4-20mA Analog Output Scale for Flow

28 FlowOutOffset_mA - DAC 4-20mA Analog Output Offset for Flow

GAS TEMPERATURE

29 TempOutScaleV

- DAC 0-5/0-10 Analog Output Scale for Temperature

30 TempOutScale_mA - DAC 4-20mA Analog Output Scale for Temperature

31 TempOutOffset_mA - DAC 4-20mA Analog Output Offset for Temperature

GAS PRESSURE

32 PresOutScaleV

- DAC 0-5/0-10 Analog Output Scale for Pressure

33 PresOutScale_mA- DAC 4-20mA Analog Output Scale for Pressure

Note: Models DFM 26/36/46 do not support temperature and pressure
measurement and have only one gas flow analog output.

7.3.1 Initial Setup

Power up the Mass Flow Meter for at least 3 minutes prior to commencing the cal-
ibration procedure. Make sure absolutely no flow takes place through the meter.
Establish digital RS-485/RS-232 communication between PC (communication ter-
minal) and DFM. The commands provided below assume that calibration will be
performed manually (w/o Aalborg

7 supplied calibration and maintenance soft-

ware) and the device has RS-485 address 11. If Aalborg

7 supplied calibration and

maintenance software is used, skip the next section and follow prompts from the
software.

Enter Backdoor mode by typing: !11,MW,1000,1[CR]
Unit will respond with:

!11,BackDoorEnabled: Y

Disable DAC update by typing:

!11,WRITE,4,D[CR]

Unit will respond with:

!11,DisableUpdate: D

,