Aalborg DFM Digital User Manual

Page 40

Advertising
background image

36

5.

Save TempOutScaleV in to the EEPROM: !11,MW,29,X[CR]

Where: X - the calculated TempOutScaleV value.

7.3.5 Gas temperature 4-20 mA analog output

calibration (DFM 27/37/47)

1. Install jumpers J3 on the analog PC board for 4-20 mA output (see Table VI).
2.

Connect a certified high sensitivity multi-meter set for the current
measurement to the pins 3 (+) and 16 (-) of the 25 pins D connector.

3.

Write 4000 counts to the DAC channel 2: !11,WRITE,2,4000[CR]

4.

Read current with the meter and calculate:

5.

Write zero counts to the DAC channel 2: !11,WRITE,2,0CR]

6.

Read offset current with the meter and calculate:

7. Save TempOutScale_mA in to the EEPROM:

!11,MW,30,Y[CR]

Save TempOutOffset_mA in to the EEPROM:

!11,MW,31,Z[CR]

Where: Y - the calculated TempOutScale_mA value.

Z - the calculated TempOutOffset_mA value.

7.3.6 Gas pressure 0-5 Vdc analog output

calibration (DFM 27/37/47)

1. Install jumpers J4 on the analog PC board for 0-5 Vdc output (see Table VI).
2.

Connect a certified high sensitivity multi-meter set for the voltage
measurement to the pins 4 (+) and 17 (-) of the 25 pin D connector.

3.

Write 4000 counts to the DAC channel 3: !11,WRITE,3,4000[CR]

4.

Read voltage with the meter and calculate:

5.

Save PresOutScaleV in to the EEPROM: !11,MW,32,X[CR]
Where: X - the calculated PresOutScaleV value.