50

APPENDIX I

AALBORG

7 DFM EEPROM Variables Rev.A3 [12/05/2004]

Gas Independent Variables

INDEX

NAME

DATA TYPE

NOTES

0

BlankEEPROM

char[10]

Do not modify. Table Revision.

1

SerialNumber

char[20]

Serial Number

2

ModelNumber

char[20]

Model Number

3

SoftwareVer

char[10]

Firmware Version

4

TimeSinceCalHr

float

Time since last calibration in hours.

5

Options1

uint

Misc. Options*

6

Reserved2

char

Flow Conditions [0-STD, 1-ACTUAL]

7

AddressRS485

char[3]

Two character address for RS485 only

8

DigGasNumber

int

Current Gas Table Number [0-9]

9

FlowUnits

int

Current Units of Measure [0-10]

10

AlarmMode

char

Alarm Mode ['R’- Enabled, 'S’ - Disabled]

11

LowAlarmPFS

float

Low Flow Alarm Setting [%FS] 0-Disabled

12

HiAlarmPFS

float

High Flow Alarm Setting [%FS] 0-Disabled

13

AlmDelay

uint

Flow Alarm Action Delay [0-3600sec] 0-Disabled

14

RelaySetting[0]

char

Relay #1 Assignment Setting (N, T, H, L, Range)

15

RelaySetting[1]

char

Relay #2 Assignment Setting (N, T, H, L, Range)

16

TotalMode

char

Totalizer Mode ['R’- Enabled, 'S’ - Disabled]

17

Total

float

Totalizer Volume in %*s (updated every 6 min)

18

TotalFlowStart

float

Start Totalizer at flow [%FS] 0 - Disabled

19

TotalVolStop

float

Totalizer Action Limit Volume [%*s] 0-Disabled

20

KfactorIndex

int

Internal K-Factor Index [0-31]**

21

UserDefKfactor

float

User Defined K-Factor

22

UDUnitKfactor

float

K-Factor for User Defined Units of Measure

23

UDUnitTimeBase

int

User Defined Unit Time Base [1, 60, 3600 sec]

24

UDUnitDensity

char

User Defined Unit Density Flag [Y, N]

25

TPUnits

char

Temperature/Pressure Units [C (Bar), F (PSI)]

26

FlowOutScaleV

float

DAC 0-5/0-10 Analog Output Scale for Flow

27

FlowOutScale_mA

float

DAC 4-20mA Analog Output Scale for Flow

28

FlowOutOffset_mA

float

DAC 4-20mA Analog Output Offset for Flow

29

TempOutScaleV

float

DAC 0-5/0-10 Analog Output Scale for Temp.

30

TempOutScale_mA

float

DAC 4-20mA Analog Output Scale for Temp.

31

TempOutOffset_mA

float

DAC 4-20mA Analog Output Offset for Temp.

32

PresOutScaleV

float

DAC 0-5/0-10 Analog Output Scale for Pressure

33

PresOutScale_mA

float

DAC 4-20mA Analog Output Scale for Pressure

34

PresOutOffset_mA

float

DAC 4-20mA Analog Output Offset for Pressure

Advertising