7.3.7 Gas pressure 4-20 mA analog output

calibration (DFM 27/37/47)

1. Install jumpers J4 on the analog PC board for 4-20 mA output (see Table VI).
2.

Connect a certified high sensitivity multi-meter set for the current
measurement to the pins 4 (+) and 17 (-) of the 25 pin D connector.

3.

Write 4000 counts to the DAC channel 3: !11,WRITE,3,4000[CR]

4.

Read current with the meter and calculate:

5.

Write zero counts to the DAC channel 3: !11,WRITE,3,0CR]

6.

Read offset current with the meter and calculate:

7. Save PresOutScale_mA in to the EEPROM:

!11,MW,33,Y[CR]

Save PresOutOffset_mA in to the EEPROM:

!11,MW,34,Z[CR]

Where: Y - the calculated PresOutScale_mA value.

Z - the calculated PresOutOffset_mA value.

Note: When done with the analog output calibration make sure the DAC
update is enabled and the BackDoor is closed (see command below).

Enable DAC update by typing:

!11,WRITE,4,N[CR]

Unit will respond with:

!11,DisableUpdate: N

Close Backdoor mode by typing:

!11,MW,1000,0[CR]

Unit will respond with:

!11,BackDoorEnabled: N

7.4

Temperature or/and Pressure sensor
Calibration

Calibration of the temperature and pressure sensors for DFM 27/37/47 devices is
not described in this manual. Temperature or/and pressure sensors re-calibration
requires factory assistance.

8.

RS-485/RS-232 SOFTWARE INTERFACE
COMMANDS

8.1 General

The standard DFM comes with an RS-485 interface. For the optional RS-232 inter-
face the start character (!) and two hexadecimal characters for the address have
to be omitted. The protocol described below allows for communications with the

,