7.3.2 Gas flow 0-5 Vdc analog output calibration

1.

Install jumpers J2 on the analog PC board for 0-5 Vdc output (see Table VI).

2.

Connect a certified high sensitivity multi meter set for the voltage
measurement to the pins 2 (+) and 15 (-) of the 25 pins D connector.

3.

Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]

4.

Read voltage with the meter and calculate:

5.

Save FlowOutScaleV in to the EEPROM: !11,MW,26,X[CR]
Where: X - the calculated FlowOutScaleV value.

7.3.3 Gas flow 4-20 mA analog output calibration

1. Install jumpers J2 on the analog PC board for 4-20 mA output (see Table VI).
2.

Connect a certified high sensitivity multi meter set for the current
measurement to pins 2 (+) and 15 (-) of the 25 pins D connector.

3.

Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]

4.

Read current with the meter and calculate:

5.

Write zero counts to the DAC channel 1: !11,WRITE,1,0CR]

6.

Read offset current with the meter and calculate:

7. Save FlowOutScale_mA in to the EEPROM:

!11,MW,27,Y[CR]

Save FlowOutOffset_mA in to the EEPROM:

!11,MW,28,Z[CR]

Where:

Y - the calculated FlowOutScale_mA value.
Z - the calculated FlowOutOffset_mA value.

7.3.4 Gas temperature 0-5 Vdc analog output

calibration (DFM 27/37/47)

1. Install jumpers J3 on the analog PC board for 0-5 Vdc output (see Table VI).
2.

Connect a certified high sensitivity multi-meter set for the voltage
measurement to the pins 3 (+) and 16 (-) of the 25 pin D connector.

3.

Write 4000 counts to the DAC channel 2: !11,WRITE,2,4000[CR]

4.

Read voltage with the meter and calculate: