1. Press the SET key.

2. Press the START DELAY key.

3. Press an ARROW key. Continue until
Level Control is shown on the display.

4. Press the ENTER key.

The display will change to read “Minimum
Contact Time: 03 seconds”. The actual time
shown may be any two digit number. The
first digit will be flashing as a prompt for
input.

Press a number key (0-9). The number will
replace the current number and advance the
flashing prompt to the next digit. Press a
second number key (0-9). Repeat this
procedure if number is wrong, until desired
time is displayed. Press the ENTER key.

The display will echo the last entry with
START DELAY, <ENTERED>.

The controller has now been given a new
value for START DELAY. The new
values reside in the NEW ENTRIES area of
the controller memory. To make these
changes active, press the RESTART key;
press it again to confirm your choice. The
controller will then wait until it receives a
contact closure (at the External Start
inputs on the terminal block). The contact
must remain closed for the length of time
programmed in the steps above.